|
As with any design,
several iterative testing steps were necessary to test and
verify the functionality of our proposed design. Below
is the approximate progression of testing and verification:
I. Preliminary Testing
After receiving the data
acquisition and accelerometers, preliminary testing was
necessary to determine whether the two separate systems were
compatible. As seen in the first picture to the right, a
cellular telephone set to vibrate ring mode was utilized to
simulate a relatively high frequency similar to that found of
the high frequency ventilators.
II. Shaker Testing
The mechanical
engineering department vibration laboratory facilities were
used to conduct some initial analysis as to what filtering
techniques were necessary. Following advice given by the
accelerometer manufacturer, a low-pass filter was constructed.
A voltage regulator was also introduced to maintain a
consistent 5 Volt supply to the accelerometer via two 6 V
batteries in series. The initial breadboard setup can be
seen in the second picture to the right.
III. Hershey Plastic Test Lung
Realistic testing was
necessary to encompass any variation in lung compliance and
background noise from the ventilator. Several test
scenarios were used to simplify and pinpoint sources of error
which will be considered for the final deliverable prototype.
The plastic lung also enabled the team to vary the flow into
each of the two billows as seen in the third picture to the
right.
IV. Neonatal Patient Testing
Upon successful
completion of the design project, the Neonatal PALL team
intends to conduct testing on patient at Hershey's NICU.
One such patient on a high frequency ventilator during one of
team's trips to Hershey can be seen in the fourth picture to
the right. Per Dr. Palmer's request, global changes will
be recorded for significant time periods to help determine
global changes in ventilator effectiveness and will be
displayed in two displays: one showing real-time results and a
second display to general trends. |